| Catalog Number |
0165L18 |
| Medical Device Problem Code |
Material Puncture/Hole (1504)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/18/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that product had a hole in foley catheter bifurcation.(pcn#(b)(4), pcn#0165l18).
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was inconclusive because this investigation did not result in any additional findings and no sample was available for evaluation.The reported event is addressed within the labeling as it states: warning: on catheter, do not use ointments or lubricants having petrolatum base.They will damage latex and may cause the balloon to burst.Caution: do not aspirate urine through drainage funnel wall.Recommended inflation capacities: 5cc balloon: use 10ml sterile water.Do not exceed recommended capacities.Visually inspect the product for any imperfections or surface deterioration prior to use.A dhr could not be performed since no lot number was provided.No additional actions can be taken at this time.Correction: d.Udi format updated with available product information per gudid.H11: sections a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that product had a hole in foley catheter bifurcation.(pcn#165818, pcn#0165l18).Per additional information received via email on 25sep2025, it was reported that the product was already returned.
|
| |
|
Search Alerts/Recalls
|