MEDTRONIC, INC. CAPSURE EPICARDIAL LEAD; DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) Back to Search Results
Model Number 4965-25
Medical Device Problem Codes Failure to Capture (1081); High impedance (1291); Low impedance (2285)
Health Effect - Clinical Code Syncope/Fainting (4411)
Date of Event 08/23/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Continuation of d10:  w1dr01 ipg implanted on (b)(6) 2025.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that the right atrial (ra) lead exhibited out of range and low impedance warning in unipolar configuration and undefined and high impedance in bipolar configuration.The right ventricular (rv) lead exhibited low impedance, atrial unipolar lead was also programmed bipolar, leading to lead issues and hospitalization for loss of capture.The patient experienced syncope.The leads were capped and replaced.No further patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPSURE EPICARDIAL LEAD
Common Device Name
DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV)
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112-4391
7635269277
MDR Report Key23015998
Report Number2182208-2025-04525
Device Sequence Number10038317
Product Code NVN
UDI-Device Identifier00613994501745
UDI-Public(01)00613994501745(17)220214
Combination Product (Y/N)Y
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P950024
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date02/14/2022
Device Model Number4965-25
Device Catalogue Number4965-25
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Fifteen-Day
Initial Date Received by Manufacturer 08/23/2025
Initial Report FDA Received Date09/10/2025
Date Device Manufactured02/25/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
4968-25 LEAD
Outcome Attributed to Adverse Event Life Threatening; Required Intervention; Hospitalization;
Patient Age5 YR
Patient SexMale
Patient Weight19 KG
-
-