| Model Number |
4965-25 |
| Medical Device Problem Codes |
Failure to Capture (1081); High impedance (1291); Low impedance (2285)
|
| Health Effect - Clinical Code |
Syncope/Fainting (4411)
|
| Date of Event |
08/23/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Continuation of d10: w1dr01 ipg implanted on (b)(6) 2025.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that the right atrial (ra) lead exhibited out of range and low impedance warning in unipolar configuration and undefined and high impedance in bipolar configuration.The right ventricular (rv) lead exhibited low impedance, atrial unipolar lead was also programmed bipolar, leading to lead issues and hospitalization for loss of capture.The patient experienced syncope.The leads were capped and replaced.No further patient complications have been reported as a result of this event.
|
| |
|