| Catalog Number |
IAB-S730C |
| Medical Device Problem Code |
Inflation Problem (1310)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/13/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported "after the catheter inserted, the balloon can not inflate completely.Change the new catheter".No patient injury or consequence reported.The patient's current condition is reported as "fine".
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed, and no relevant findings were identified.Without the device to evaluate, the complaint could not be confirmed, and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
|
| |
|
Event or Problem Description
|
|
It was reported "after the catheter inserted, the balloon can not inflate completely.Change the new catheter".No patient injury or consequence reported.The patient's current condition is reported as "fine".
|
| |
|
Event or Problem Description
|
|
It was reported "after the catheter inserted, the balloon can not inflate completely.Change the new catheter".No patient injury or consequence reported.The patient's current condition is reported as "fine".
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).The serial number on the returned sample is (b)(6).The lot number for the returned sample is (b)(6).Additional information obtained from a teleflex representative indicates the returned sample is the correct iabc for this complaint.Returned for investigation was a 40cc 7.0fr rediguard intra-aortic balloon catheter (iabc) with the original packaging box that does not match the serial number of the returned sample.The sample was returned in a cardboard box and was loosely packed within the original packaging tray/carton.Upon return, the one-way valve was con-nected and tethered to the short driveline tubing.The bladder was fully unwrapped.Dried blood was noted on the exterior surfaces of the returned sample.No obvious blood was noted within the helium pathway.No visual damage or abnormalities were noted to the returned sample.The bladder thick-ness was measured at six points with measurements ranging from 0.0062n-0.0068in and was within specification of process document.The one-way valve was tested and passed.A vacuum was pulled on the one-way valve, and it held for at least 1 minute according to quality system document.An attempt to aspirate and flush the catheter using a 60cc lab-inventory syringe was unable to be suc-cessfully completed due to a blocked/clotted central lumen.Im mediate push back on the syringe plunger was experienced.The blockage, most likely dried blood, was unable to be cleared.The iabc was connected to an iabp with a lab inventory 30cc inflation driveline tubing.The iabc was inserted into the "t" tube and 75mmhg backpressure was applied.The iabc was pumped for a minimum of 15 minutes.There were no alarms triggered.The bladder inflated and deflated completely with each beat.The iabc was leak tested in accordance with testing methods from manufacturing procedure.No leaks were detected.Full inflation was achieved.The device passed the leak test.A lab inventory 0.025in guidewire was back loaded through the iabc distal tip.The guidewire could not advance at approximately 56.1cm from the iabc distal tip, which is the location of the clotted central lumen.Dried blood was noted on the guidewire upon removal.The guidewire was front loaded through the iabc luer.The guidewire could not advance at approximately 18.9cm from the iabc luer, which is the location of the clotted central lumen.Dried blood was noted on the guidewire upon removal.A device history record (dhr) review was conducted for the lot number with no re levant findings.The device passed all manufacturing specifications prior to release.The reported complaint that the "balloon cannot inflate completely" is not confirmed.The returned iabc bladder was fully intact with no abnormalities noted.During the investigation, the iabc fully inflated, and no leaks/alarms were noted.The returned device passed visual and functional test specifications.Based on a review of the device history record (dhr), the product met specification upon release.The root cause of the complaint is undetermined.No further action required at this time.
|
| |
|
Search Alerts/Recalls
|