| Catalog Number |
309680 |
| Medical Device Problem Code |
Contamination /Decontamination Problem (2895)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/03/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H.3.If a device evaluation and/or device history review is completed, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that bd syringe 50cc ll tip convenience pak - visible droplets.It was reported by customer that they found to have liquid droplets in the rubber plunger ridges.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation results: our quality engineer inspected the 5 photos and 11 physical samples submitted for evaluation.The reported issue of visible droplets was not confirmed upon inspection of the samples.Analysis of the sample showed that visual inspection under magnification showed no defects were observed and each sample passed a leak test.Bd could not determine a manufacturing related root cause since the reported defect was not confirmed during the evaluation of the sample.A review of our risk management documentation was performed and indicates that the potential risk of the reported event was assessed appropriately.Production records were reviewed, and this batch meets our manufacturing specification requirements.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
No additional information.
|
| |
|