| Catalog Number |
0119L18 |
| Medical Device Problem Codes |
Material Fragmentation (1261); Material Puncture/Hole (1504)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/04/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Initial reporter name: (b)(6) hospital.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that before use, foley catheter tip was broken.Per follow up information received on 09sep2025, customer confirmed that there was no patient involvement.
|
| |
|
Event or Problem Description
|
|
It was reported that before use, foley catheter tip was broken.Per follow up information received on 09sep2025, customer confirmed that there was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The complete initial reporter's facility name is (b)(6) hospital, (b)(6).The reported event was confirmed ¿ manufacturing related.The product had caused the reported failure.Sample was evaluated and observed there was a pinhole on the bottom part of sac collar that allowed water to leak out.A potential root cause for this failure could be due to bubble during dipping process.A review of the device labeling is adequate to prevent the reported event.A review of the manufacturing process indicates that the process is capable of producing of this type of defect.However, current in-process controls per pfmea are adequate to identify the defect or verify the product integrity.Corrections: d, f, h.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: sections a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|