|
On 08/21/2025, a distributor reported via email that an ar-3638h knotless hip fibertak suture anchor experienced an issue in which the repair suture breaks from the anchor during passage.The case was completed with another new implant.This was discovered during a hip instability procedure on (b)(6) 2025, with no reported patient harm.Additional information has been requested on 08/27/2025.
|
|
Additional information: g3, h3, h6.Complaint allegation is not confirmed.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.Per dfu-0054-eo - e.Warnings - 6.Preoperative and operating procedures, including knowledge of surgical techniques and proper selection and placement of the device, are important considerations in the successful utilization of this device.The appropriate arthrex delivery system is required for proper implantation of the device.7.Avoid excessive impaction during anchor insertion as this could lead to inserter damage and/or breakage.If insertion resistance is encountered, do not impact harder.Replace the implant and repeat the drilling/insertion process.G.Precautions - 6.Suturetak and fibertak suture anchors only: anchors loaded on flexible drivers: the drill guide tip must remain in contact with the bone surface during the drilling and implant impaction steps of the procedure.Failure to do so may result in difficulty seating the implant to its intended depth.Avoid excessive impaction during anchor insertion as this could lead to inserter damage and/or breakage.If insertion resistance is encountered, do not impact harder.Replace the implant and repeat the drilling/insertion process.7.Arthrex implant specific instrumentation must be used to insert implantable devices per the recommended surgical technique.The method of bone preparation and the quality of the bone encountered were not specified.The most likely cause of the reported failure can be attributed to user error of the device due to improper bone preparation, or wrong surgical technique during the knotless mechanism conversion process, the device contact with another instrument during use, and/or excessive force applied during suture passage, which could compromise suture integrity.
|