The reported event was confirmed since images of ct scans were provided and shows loosening and subsidence of the tibial and talar components.Upon further investigation of the ct scans by healthcare professionals the following was observed: the ct scan shows radiolucence and some cysts adjacent to the tibial component, suggesting loosening and anterior subsidence, therefore migration.There is also some radiolucence and a slight anterolateral subsidence visible for the talar component, which would suggest some loosening and migration here as well.There is no valuable clinical information given in this case, therefore apart from the loosening and the subsidence of both components, no assessment regarding the underlying cause of the revision can be made.However, failure of total ankle replacement tar over time is a known complication.In case of a planned revision procedure a medical opinion aiming to determine the root cause of the failure is part of the internal investigation process.Sufficient radiological and clinical information must be provided to enable a meaningful clinical assessment and to identify possible causes of failure.In cases where a ct scan is the only available clinical source, this assessment is limited.No conclusive statement can be provided, because key clinical information e.G., clinical status, exact symptoms, and range of motion of the patient, assessment of the treating physician is missing.The root causes of radiographic findings such as radiolucent areas and bone cysts are often complex and multifactorial and cannot be determined scientifically without this decisive evidence.Due to these limitations, we are unable in such cases to provide statements about the patient, the procedure and or the device in relation to cause of the failure.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If the device is returned or if any additional information is provided, the investigation will be reassessed.
|