• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY CMW - 9610921 DEPUY CMW 2 20G; BONE CEMENT : BONE CEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY CMW - 9610921 DEPUY CMW 2 20G; BONE CEMENT : BONE CEMENT Back to Search Results
Catalog Number 3322020
Medical Device Problem Codes Device Damaged Prior to Receipt by User (1570); Device Damaged Prior to Use (2284)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/25/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that patient underwent total knee arthroplasty (tka) on (b)(6) 2025.During the procedure, two fast set bone cement vials from the same lot exploded into fragments with sharp edges when opened.All fragments were removed.A new cement was opened away from the patient and other instrumentation.Procedure was completed successfully with a delay of five minutes.There was no patient consequence.
 
Additional Manufacturer Narrative
Product complaint # (b)(4) this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees hat the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H11 additional narrative: added: d10.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint #(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees hat the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H11 additional narrative: added: h6 (health effect - impact code).Corrected: h6 (medical device problem code).If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H11 additional narrative: corrected: d3, h6 (health effect - impact code).If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary.No device associated with this report was received for examination.A manufacturing record evaluation was performed for the finished device [3322020 / 4600707] number, and no non-conformances / manufacturing irregularities related to the complaint condition were identified.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot.A manufacturing record evaluation was performed for the finished device [3322020 / 4600707] number, and no non-conformances / manufacturing irregularities related to the complaint condition were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEPUY CMW 2 20G
Common Device Name
BONE CEMENT : BONE CEMENT
Manufacturer (Section D)
DEPUY CMW - 9610921
cornford rd
cornford rd
blackpool IN FY4 4 QQ
UK  FY4 4QQ
Manufacturer (Section G)
DEPUY CMW - 9610921
cornford rd
blackpool FY4 4 QQ
UK   FY4 4QQ
Manufacturer Contact
elba bello
700 orthopaedic dr.
warsaw, IN 46581
9083863534
MDR Report Key23019378
Report Number1818910-2025-15628
Device Sequence Number13203741
Product Code LOD
UDI-Device Identifier10603295168621
UDI-Public(01)10603295168621(17)270930(10)4600707
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K053003
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number3322020
Device Lot Number4600707
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 08/25/2025
Supplement Date Received by Manufacturer09/09/2025
09/29/2025
10/07/2025
11/14/2025
Initial Report FDA Received Date09/10/2025
Supplement Report FDA Received Date09/12/2025
10/01/2025
10/08/2025
11/14/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DEPUY CMW 2 20G.
Patient SexUnknown
-
-