Catalog Number 305540
Medical Device Problem Code Material Perforation (2205)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/27/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
B.3.The date received by manufacturer has been used for this field h.3.If a device evaluation and/or device history review is completed, a supplemental report will be filed.
 
Event or Problem Description
It was reported that the bd syringe allergy 1ml w/ndl 27x1/2 rb needle went through the shield.The following information was provided by the initial reporter: material # 305540 , batch # 5097095.Verbatim: rcc received a complaint via email.Email(s) attached.I wasn't sure if i let you know or if you know who i reached out to but i had an incident in the past that has now happened again.It might be a defect in the order but i wanted to bring it to someone attention since it's the second time.One of our needles was sticking out of the top, i have attached pictures for reference.I¿m not sure if this is a business unit that you can help with but i need someone to get in touch with regarding issues my customer is having.This isn¿t the first incident they¿ve experienced with broken needles from the bd allergy trays.I¿ve included the photos they sent me.If you¿re able to help me get them working product that would be great, or get me in contact with who can help! the customer experienced with broken needles from the bd allergy trays.Item# 305540.Lot# 5097095.
 
Additional Manufacturer Narrative
2 photos were received and analyzed by our quality team with the customer's complaint of needle through shield.The photos are enough to verify the failure but without a physical sample for investigation, the root cause remains unknown.Multiple attempts were made to acquire the device production history but due to plant closure the records could not be obtained.If there were quality alerts or abnormalities for this lot# these would be evident and reported.In this case quality defects in this lot are unlikely.
 
Event or Problem Description
Photos.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD SYRINGE ALLERGY 1ML W/NDL 27X1/2 RB
Common Device Name
PISTON SYRINGE
Manufacturer (Section D)
EMBECTA MEDICAL I LLC
1329 west highway 6
holdrege NE 68949
Manufacturer (Section G)
EMBECTA MEDICAL I LLC
1329 west highway 6
holdrege NE 68949
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key23020343
Report Number2243072-2025-01145
Device Sequence Number15839366
Product Code FMF
UDI-Device Identifier00382903055401
UDI-Public(01)00382903055401(17)300430(10)5097095
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K110771
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/16/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number305540
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/27/2025
Supplement Date Received by Manufacturer12/16/2025
Initial Report FDA Received Date09/10/2025
Supplement Report FDA Received Date01/14/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-