• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE PKG ASSY. PEEK SOFT EYELET RC ANCHOR 5.5MM; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE PKG ASSY. PEEK SOFT EYELET RC ANCHOR 5.5MM; SCREW, FIXATION, BONE Back to Search Results
Catalog Number 3910-400-010
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/11/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.
 
Event or Problem Description
It was reported that there was breakage of peek part of peek intraline anchor post drilling and tapping.All pieces were retrieved.
 
Event or Problem Description
It was reported that there was breakage of peek part of peek intraline anchor post drilling and tapping.All pieces were retrieved.
 
Additional Manufacturer Narrative
Three attempts have been made to retrieve the device for evaluation, and it is yet to be received in-house.This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: anchor breaking.Probable root cause: application.Excessive force torquing anchor or lateral force applied during insertion.Not enough axial force during insertion.Use of wrong tap, improperly prepared hole.Bone to hard for anchor insertion.Bone not hard enough to maintain screw purchase.No pilot hole prepared.The reported failure mode will be monitored for future reoccurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG ASSY. PEEK SOFT EYELET RC ANCHOR 5.5MM
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
arsalan rizvi
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key23022286
Report Number0002936485-2025-00772
Device Sequence Number13148168
Product Code HWC
UDI-Device Identifier07613154644568
UDI-Public07613154644568
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K071157
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/11/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number3910-400-010
Device Lot Number24234AE2
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/13/2025
Supplement Date Received by Manufacturer12/04/2025
Initial Report FDA Received Date09/10/2025
Supplement Report FDA Received Date12/11/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/21/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age53 YR
Patient SexFemale
-
-