RESPIRONICS, INC. SIMPLYGO MINI; GENERATOR, OXYGEN, PORTABLE Back to Search Results
Model Number 1113604
Medical Device Problem Code Overheating of Device (1437)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/18/2025
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer received a complaint alleging that the simplygo mini device had no battery charge.The device was returned to the third-party service center for evaluation, and the customer¿s complaint was confirmed.During the evaluation, it was noted that the device's main pca was burned.Additionally, during the evaluation of the device, the field service engineer observed that the sieves were clogged, the compressor was contaminated, and the air side & o2 side were defective.The device will be returned unrepaired to the manufacturer's product investigation lab for further investigation.The root cause is not confirmed at this time.The device has not yet been returned to the manufacturer for evaluation.If any additional information is received, a follow-up report will be filed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIMPLYGO MINI
Common Device Name
GENERATOR, OXYGEN, PORTABLE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
7247330200
MDR Report Key23023074
Report Number2518422-2025-109962
Device Sequence Number18514419
Product Code CAW
UDI-Device Identifier00606959032422
UDI-Public(01)00606959032422
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K111885
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/11/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number1113604
Device Catalogue Number1113604
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/30/2025
Initial Report FDA Received Date09/10/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/05/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-