| Catalog Number |
68-800 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Healthcare professional reported "side exchange with no complaint against the device".Later healthcare professional reported "deflation" this record relates to the left side.Device was explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: deflation.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the product analysis performed, the assessments of the complaint codes are: ¿ deflation: observed an opening assessed as fold flaw opening.As per the investigation procedure, wear abrasion and creases were observed and none of the observations are found to be potentially related to the manufacturing process, no further actions are required.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "side exchange with no complaint against the device".Later healthcare professional reported "deflation" this record relates to the left side.Device was explanted and replaced.Device has been returned.
|
| |
|