Catalog Number P3122010
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/08/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The baxter technician replaced the power supply to resolve the issue.Based on this information, no further actions are required at this time.Likorall overhead lift is intended for use in, health care, intensive care and rehabilitation.Likorall overhead lift is designed for fixed installation and free-standing lift systems.All common lifts and transfers can be performed using likorall overhead lift, for instance between bed/wheelchair, to/from floor, toilet visits, gait training, and together with stretchers.Likorall r2r (room to room) overhead lift enables the patient to be moved between two rail systems in separate rooms.Likorall overhead lift with the es designation is prepared for operation with the wireless handcontrol remote (ir) and in addition, a transfer motor can be connected for motor driven movement along the rail.Likorall s, irc overhead lift is prepared for continuous charging through the railsystem.Although there was no adverse event reported, if the reported malfunction were to recur, it could result in serious injury or death as contact with exposed electrical wire can result in electrical injury.
 
Event or Problem Description
During servicing by baxter, technical service identified that the likorall 242 s, charger was damaged with exposed copper.This occurred during baxter servicing/testing.There was no patient/user injury, medical intervention, symptom, or adverse event reported.There were no other devices reported to be involved.The customer did not communicate this event to another baxter department or authority.The device was requested for service/inspection by baxter.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIKORALL 250
Common Device Name
LIFT PATIENT NON AC POWERED
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
LIKO AB
nedre vagen 100
lulea, norrbottens lan [se-25] 975 9 2
SW   975 92
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key23027647
Report Number8030916-2025-00107
Device Sequence Number9514173
Product Code FSA
UDI-Device Identifier00887761979819
UDI-Public(01)00887761979819
Combination Product (Y/N)N
Initial Reporter CountryFR
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 09/11/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberP3122010
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/03/2025
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/03/2025
Initial Report FDA Received Date09/11/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
N/A.
Patient SexUnknown
-
-