The baxter technician replaced the power supply to resolve the issue.Based on this information, no further actions are required at this time.Likorall overhead lift is intended for use in, health care, intensive care and rehabilitation.Likorall overhead lift is designed for fixed installation and free-standing lift systems.All common lifts and transfers can be performed using likorall overhead lift, for instance between bed/wheelchair, to/from floor, toilet visits, gait training, and together with stretchers.Likorall r2r (room to room) overhead lift enables the patient to be moved between two rail systems in separate rooms.Likorall overhead lift with the es designation is prepared for operation with the wireless handcontrol remote (ir) and in addition, a transfer motor can be connected for motor driven movement along the rail.Likorall s, irc overhead lift is prepared for continuous charging through the railsystem.Although there was no adverse event reported, if the reported malfunction were to recur, it could result in serious injury or death as contact with exposed electrical wire can result in electrical injury.
|
|
During servicing by baxter, technical service identified that the likorall 242 s, charger was damaged with exposed copper.This occurred during baxter servicing/testing.There was no patient/user injury, medical intervention, symptom, or adverse event reported.There were no other devices reported to be involved.The customer did not communicate this event to another baxter department or authority.The device was requested for service/inspection by baxter.
|