• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LUTRONIC CORPORATION XERF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LUTRONIC CORPORATION XERF Back to Search Results
Catalog Number F1101300330
Medical Device Problem Code Defective Component (2292)
Health Effect - Clinical Code Full thickness (Third Degree) Burn (2696)
Date of Event 08/13/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during corporate training, a staff member experienced a burn on the face during a xerf rf electrocoagulation and hemostasis treatment.Review of the treatment parameters used were found to be within the recommended clinical guidelines.Images provided on 08/13/2025 were reviewed and discussed with cynosure lutronic medical director.Cynosure medical director states, "the redness observed in the images suggests a partial thickness burn and the patient is at risk for scarring and/or hyper/hypopigmentation." the medical director completed a review of the additional images suppled between 08/14/2025-08/27/2025 and stated "it appears that there may be a small full thickness burn on the left jawline." a trained cynosure lutronic field service engineer (fse) evaluated the device.During this time, the device was subjected to multiple tests including full functionality testing with passing results.The device was found to be working to device published specifications.However, images of the device tip were reviewed and observed a mark which appears to be a burn mark.It is unknown what caused this visual discrepancy.At this time, it is unknown what contributed to the reported malfunction and burning observed in the patient images or the damage to the tip observed in the provided image.It is unknown what caused this observed defect to the tip.The defect was observed but unable to be verified if it contributed to the reported burning.Since a reported malfunction and serious injury occurred, we are reporting this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XERF
Common Device Name
XERF
Manufacturer (Section D)
LUTRONIC CORPORATION
lutronic center
219 sowon-ro
deogyang-gu, goyang-si 10534
KS  10534
MDR Report Key23027965
Report Number1222993-2025-00035
Device Sequence Number18505434
Product Code GEI
UDI-Device Identifier08809447657981
UDI-Public08809447657981
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Importer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 09/11/2025
Report Date (Section F) 09/11/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberF1101300330
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date08/13/2025
Event Location Other
Date Report to Manufacturer09/11/2025
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date09/11/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexFemale
Date Report Sent to FDA09/11/2025
-
-