Catalog Number MWM054
Medical Device Problem Codes Break (1069); Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/13/2025
Type of Reportable Event Malfunction
Event or Problem Description
Broken impactor tip.
 
Additional Manufacturer Narrative
Once the investigation has been completed any additional information will be reported in a supplemental report.
 
Event or Problem Description
Broken impactor tip.
 
Additional Manufacturer Narrative
Correction: please see h6 (device code).The reported event could be confirmed since the product was returned for evaluation and matches the alleged failure mode.The device inspection revealed the following: the visual inspection of the returned device exhibits rough and extensive usage.The impactor tip is broken into two pieces at the proximal end where the humeral head of impactor connects the handle.The nature of breakage pattern indicates to be a brittle fracture.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material manufacturing or design related problems were found during the investigation.Based on investigation the root cause was attributed to combination of design and normal usage related issue.The failure was caused by its intended use.As a device that is manufactured of plastic and incurs numerous impaction events, cleaning and sterilization cycles breakage as noted in this complaint can be expected.A design improvement is in progress to prevent the recurrence of the alleged failure.If any further information is provided the complaint report will be updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERFORM HUMERAL HEAD IMPACTOR TIP
Common Device Name
IMPACTOR
Manufacturer (Section D)
TORNIER INC
10801 nesbitt avenue s
bloomington MN 55437
Manufacturer (Section G)
TORNIER INC
10801 nesbitt avenue s
bloomington MN 55437
Manufacturer Contact
dannielle stap
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key23028249
Report Number0001649390-2025-00655
Device Sequence Number15907668
Product Code HWA
UDI-Device Identifier00846832082964
UDI-Public(01)00846832082964(10)CV0122146
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/24/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberMWM054
Device Lot NumberCV0122146
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/02/2025
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 08/14/2025
Supplement Date Received by Manufacturer10/30/2025
Initial Report FDA Received Date09/11/2025
Supplement Report FDA Received Date11/24/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/27/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-