• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MENTOR TEXAS MENTOR CPX 4 BREAST TISSUE EXPANDER WITH SUTURE TABS; EXPANDER, SKIN, INFLATABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MENTOR TEXAS MENTOR CPX 4 BREAST TISSUE EXPANDER WITH SUTURE TABS; EXPANDER, SKIN, INFLATABLE Back to Search Results
Catalog Number 3549212
Medical Device Problem Code Infusion or Flow Problem (2964)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/14/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.E1.Initial reporter phone: (b)(6).Reason for device explant and/or reoperation: n/a.Manufacturer¿s reference number: (b)(4).
 
Event or Problem Description
It was reported that a mentor cpx 4 breast tissue expander with suture tabs being used in a breast implantation procedure was found to be difficult to inflate during the operation.No adverse event was experienced by the patient.The surgeon used backup devices, and there were no reported patient consequences or procedural delays.
 
Additional Manufacturer Narrative
On sep 12, 2025, it was noticed the previous report erroneously omitted the following information: a photo of the device was returned for evaluation.The evaluation was completed on aug 26, 2025.Photo evaluation summary: upon visual evaluation of the image provided in the complaint, the expander was observed however, the reported filling issue cannot be confirmed based on the images provided.As the product involved in this complaint was not received, the device could not be analyzed according to our procedures.Mentor also performed a manufacturing record evaluation related to the reported complaint for the finished device lot number.No manufacturing non-conformances were identified as part of this evaluation.As part of mentor's quality process, all devices are manufactured, inspected, and released according to approved specifications.Based on the results of this investigation, it was determined that the product met the manufacturing release criteria.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR CPX 4 BREAST TISSUE EXPANDER WITH SUTURE TABS
Common Device Name
EXPANDER, SKIN, INFLATABLE
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
henry boden
3041 skyway circle north
irving, TX 75038
9497882078
MDR Report Key23031399
Report Number1645337-2025-10095
Device Sequence Number21127454
Product Code LCJ
UDI-Device Identifier00081317006725
UDI-Public(01)00081317006725(17)270816(10)9951444
Combination Product (Y/N)N
Initial Reporter CountryAR
PMA/510(K) Number
K130813
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number3549212
Device Lot Number9951444
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 08/19/2025
Supplement Date Received by Manufacturer09/12/2025
Initial Report FDA Received Date09/11/2025
Supplement Report FDA Received Date09/12/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/18/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-