MAUDE Adverse Event Report: ARTHREX, INC. ANGLED REAMER, DRIVE SHAFT; MANUAL INSTR, GENERAL SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. ANGLED REAMER, DRIVE SHAFT; MANUAL INSTR, GENERAL SURGICAL Back to Search Results
Model Number ANGLED REAMER, DRIVE SHAFT
Medical Device Problem Code Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/22/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Event or Problem Description
On 08/22/2025, it was reported by a sales representative via (b)(4) that an ar-9597-10 angled reamer sleeve and an ar-9676 angled reamer drive shaft were welded together and did not spin freely.This was discovered during the case with no patient harm.
 
Additional Manufacturer Narrative
Additional information: d9, g3, h3, h6.- one unpackaged ar-9676, angled reamer, drive shaft, batch 022406, was received for investigation.- visual evaluation of the returned device noted grind marks on both the distal and proximal end.- functional testing was performed with a known-good ar-9597-10 and resistance was felt when rotating the device within the cannulated shaft.- the most likely cause for the reported failure can be attributed to misuse due to a force perpendicular to the drive shaft being applied while the device is spinning.This applied force creates an excess of friction and heat that ultimately has the potential to cause the device to cease functioning as intended.Application of this force perpendicular to the drive shaft is not needed for the device to function, nor is it recommended that the user do this during normal clinical use.When used usually, this issue is unlikely to present itself, which suggests that the handling of the device greatly impacts whether or not it will seize.For this reason, the direct cause is considered to be device misuse.- refer to investigation photos.- complaint allegation is confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANGLED REAMER, DRIVE SHAFT
Common Device Name
MANUAL INSTR, GENERAL SURGICAL
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key23031842
Report Number1220246-2025-03819
Device Sequence Number9428085
Product Code LXH
UDI-Device Identifier00888867324954
UDI-Public(01)00888867324954(11)240214(10)022406
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 10/16/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberANGLED REAMER, DRIVE SHAFT
Device Catalogue NumberAR-9676
Device Lot Number022406
Was Device Available for Evaluation? Yes
Distributor Facility Aware Date08/22/2025
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/22/2025
Supplement Date Received by Manufacturer10/14/2025
Initial Report FDA Received Date09/11/2025
Supplement Report FDA Received Date10/16/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/14/2024
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-