| Catalog Number |
2155-1 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Current information is insufficient to permit a valid conclusion about the cause of this event.A follow up report will be sent upon completion of the device evaluation.
|
| |
|
Event or Problem Description
|
|
It was reported that the tip of pathfinder nxt universal driver tip deformed during removal of pathfinder hardware due to adjacent segment breakdown and extension of construct.There was no reported patient impact.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: "device evaluation anticipated, but not yet begun" is erroneous and no longer applies.The device was evaluated.Corrections in h3.Additional information in h6: investigation type, findings, and conclusions.Device evaluation: the returned device matches the information in the complaint file and was examined.Visual inspection revealed one of the prongs on the tip of the device is fractured.Root cause: this event could be attributed to excessive forces applied during use, possibly due to hard patient bone.Dhr review: the dhr was reviewed.There are no indications of manufacturing issues which would have contributed to this event and the device was likely conforming when it left highridge medical¿s control.Device usage: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
|
| |
|
Event or Problem Description
|
|
It was reported that the tip of pathfinder nxt universal driver tip deformed during removal of pathfinder hardware due to adjacent segment breakdown and extension of construct.There was no reported patient impact.
|
| |
|