| Catalog Number |
07.01805.001 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/17/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Current information is insufficient to permit a valid conclusion about the cause of this event.A follow up report will be sent upon completion of the device evaluation.
|
| |
|
Event or Problem Description
|
|
It was reported that during a hardware removal surgery following a successful fusion, the tip of a rigid final driver broke off while removing the right closure top of the occipital plate.It was confirmed that no remnants were left inside the patient's body, and the procedure was completed successfully without patient impact.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: "device evaluation anticipated, but not yet begun" is erroneous and no longer applies.The device was evaluated.Corrections in h3.Additional information in h6: investigation type, findings, and conclusions.Device evaluation: the returned device matches the information in the complaint file and was examined.Visual inspection revealed the tip is twisted and fractured.Root cause: this event could possibly be attributed to off-axis or excessive forces applied during use.Dhr review: the dhr was reviewed.There are no indications of manufacturing issues which would have contributed to this event and the device was likely conforming when it left highridge medical¿s control.Device usage: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
|
| |
|
Event or Problem Description
|
|
It was reported that during a hardware removal surgery following a successful fusion, the tip of a rigid final driver broke off while removing the right closure top of the occipital plate.It was confirmed that no remnants were left inside the patient's body, and the procedure was completed successfully without patient impact.
|
| |
|