• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. SIMPLYGO; GENERATOR, OXYGEN, PORTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. SIMPLYGO; GENERATOR, OXYGEN, PORTABLE Back to Search Results
Model Number 1069058
Medical Device Problem Codes Thermal Decomposition of Device (1071); Overheating of Device (1437)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/24/2025
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer received information alleging a simplygo device had a power connection problem and the circuit board was defective.The device was not in use on a patient at the time of the occurrence.The device was returned to the third-party service center for evaluation, and the customer¿s complaint was confirmed.During the evaluation, it was noted that the device pca was burned behind the ac power connector.In addition, the unit starts with a battery as a power source, the sieve canisters were worn out, the tubing was contaminated, and the inlet filter needed to be replaced due to preventive maintenance.The device will be returned to the pil for additional testing.The root cause is not confirmed at this time.The device has not yet been returned to the manufacturer for evaluation.If any additional information is received, a follow-up report will be filed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIMPLYGO
Common Device Name
GENERATOR, OXYGEN, PORTABLE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
Respironics, Inc.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
+1(724)733-0200
MDR Report Key23032482
Report Number2518422-2025-110035
Device Sequence Number18567373
Product Code CAW
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K111885
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation 505
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number1069058
Device Catalogue Number1069058
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/28/2025
Supplement Date Received by ManufacturerNot provided
09/28/2026
Initial Report FDA Received Date09/11/2025
Supplement Report FDA Received Date09/23/2026
09/29/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/12/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-