• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ATRIUM MEDICAL CORPORATION EXPRESS; BOTTLE, COLLECTION, VACUUM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ATRIUM MEDICAL CORPORATION EXPRESS; BOTTLE, COLLECTION, VACUUM Back to Search Results
Model Number 16400
Medical Device Problem Code Off-Label Use (1494)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/06/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The most common complaints associated with express mini 500 (p/n 16400) devices are those related to outpatient and improper use of the device.In outpatient cases, the drain is set up on a patient by a clinician and then that patient is sent home with the drain where they are then responsible for its use and interpretation.This puts the responsibility of using the drain on a person who is not necessarily a trained medical professional with any knowledge of how to use and interpret the device.Device queries: the users sometimes call getinge with questions about the use and interpretation of the drain.These queries stem from the users' inexperience and lack of training with the device.The ifu states that the express mini 500 is restricted for use by trained healthcare providers familiar with cardiothoracic surgical procedures and techniques, including the use of chest drains.In addition to the instructions described for each type of complaint, the ifu states that "the express mini 500 is restricted for use in a healthcare facility.The express mini 500 should not be used for outpatient drainage." this type of complaint has been thoroughly investigated and is known to be cause by user error so further dhr review is not required to understand the issue.The ifu provides adequate instructions for the setup and use of the drain, as well as instructions about the intended users and environments.Outpatient and user error complaints of express mini 500 devices are confirmed.Device nonconformances are not confirmed for these issues.The root cause of these complaints is user error.Escalation determination: outpatient and improper use complaints of express mini 500 are attributed to user error which is currently being handled by the capa process.Trend review of these devices is completed on a monthly basis and excursions related to trending is also handled by the capa process.
 
Event or Problem Description
Home health hospice rn called from patient's home to request assistance with draining an express mini 500.The patient was sent home from an unknown facility without instructions.Rn was advised of the off-label use but given instructions for draining through the needless sampling access port, as the patient was on hospice and would not be returning to the hospital.Rn was advised to visit nearest emergency room for ongoing chest drain needs.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXPRESS
Common Device Name
BOTTLE, COLLECTION, VACUUM
Manufacturer (Section D)
ATRIUM MEDICAL CORPORATION
40 continental blvd
merrimack NH
Manufacturer (Section G)
ATRIUM MEDICAL CORPORATION
40 continental blvd
merrimack NH
Manufacturer Contact
lori gosselin
40 continental blvd
merrimack, NH 
MDR Report Key23034368
Report Number3011175548-2025-000010
Device Sequence Number9455558
Product Code KDQ
UDI-Device Identifier00650862164008
UDI-Public(01)00650862164008
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
K984496
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 09/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number16400
Device Catalogue Number16400
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/06/2025
Initial Report FDA Received Date09/11/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/20/2011
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient Age57 YR
Patient SexMale
Patient Weight90 KG
-
-