• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NAKANISHI INC. NSK; HANDPIECE, CONTRA- AND RIGHT-ANGLE ATTACHMENT, DENTAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NAKANISHI INC. NSK; HANDPIECE, CONTRA- AND RIGHT-ANGLE ATTACHMENT, DENTAL Back to Search Results
Model Number NANO95LS
Medical Device Problem Code Overheating of Device (1437)
Health Effect - Clinical Code Partial thickness (Second Degree) Burn (2694)
Date of Event 07/31/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The dentist refused to provide information about the patient's weight.
 
Event or Problem Description
On august 19, 2025, nakanishi became aware of a handpiece overheating through a complaint input into the complaint database by a distributor (nsk america).Details are as follows: - the event occurred on (b)(6) 2025.- the dentist ran the nano95ls handpiece (serial no.(b)(6)) outside of the patient's mouth to make sure water and light comes on.- within a few seconds of working on the tooth the patient reacted in pain.- the dentist asked the patient if the tooth was numb and they stated it was fine, so the dentist proceeded with the preparation of the tooth for the crown.- once the dentist removed the handpiece from the patient's mouth to change burs, the dentist noticed that the head of the handpiece was extremely hot to the touch.- this is when the dentist observed that the patient's buccal mucosa had been burned.
 
Additional Manufacturer Narrative
Upon receiving the device involved in the mdr event from the distributor, nakanishi conducted a failure analysis of the returned device, which included measuring the operating temperature of the device [report no.(b)(4)].These activities are described in more detail below.Methodology used: a) nakanishi examined the device history record and the repair history for the subject nano95ls device [serial no.(b)(6)].There were no problems observed during manufacturing or testing noted in the dhr.There were also no repair history records since the device was shipped.B) nakanishi conducted temperature testing of the returned device in the following manner: b.1) temperature sensors were attached to the exterior of the device at various test points.This included the point most proximal to the patient (testing point (1)) and points further toward the distal end of the device (testing points (2) through (4)).The test setup was prepared to take temperature measurements at all points simultaneously, including a reference measurement at ambient room temperature.B.2) nakanishi attached a thermocouple (sensor to measure temperature) to each of the testing points.Nakanishi rotated the device's motor at 40,000min-1, which is the maximum rpm for the motor that drives the handpiece (200,000min-1 for the handpiece), with water spray, and measured the exothermic response.B.3) nakanishi measured the temperature rise of the returned handpiece set at 200,000min-1 (motor revolution 40,000min-1).Nakanishi observed an abnormal temperature rise at the test points (1) and (2) about 30 seconds into the test.Temperature measurements about 30 seconds after the start of the test were as follows: test point (1): 55.0 degrees c.Test point (2): 89.5 degrees c.Test point (3): 29.9 degrees c.Test point (4): 27.2 degrees c.The increase in temperature was so sudden that the test was concluded about 30 seconds into the planned 5-mimute evaluation period.Identification of the specific failure mode(s) and/or mechanism(s) of the associated device components was conducted as follows: a) nakanishi disassembled the handpiece and performed a visual inspection of the internal parts.Nakanishi observed the following: the bearing retainer in the ball bearing on the rear side of the cartridge was broken.The cartridge was counterfeit.B) nakanishi took photographs of all the disassembled parts and kept them in the investigation report no.(b)(4).Conclusions reached based on the investigation and analysis results: a) nakanishi identified that the cause of the handpiece overheating was abnormal resistance during rotation due to the broken bearing retainer or use of the counterfeit cartridge.Nakanishi considers that the cause of the broken bearing retainer was incorporation of the counterfeit cartridge.B) misuse by the user led to the handpiece overheating., which contributed to the reported event.C) in order to prevent a recurrence of the handpiece overheating, nakanishi took the following actions: c.1) nakanishi reviewed the operation manual and reconfirmed the clarity and understandability of the instructions.C.2) nakanishi reported the above evaluation results to the distributor and directed the distributor to remind the user of the importance of using the device as instructed in the operation manual.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NSK
Common Device Name
HANDPIECE, CONTRA- AND RIGHT-ANGLE ATTACHMENT, DENTAL
Manufacturer (Section D)
NAKANISHI INC.
700 shimohinata
kanuma-shi, tochigi-ken 322-8 666
JA  322-8666
Manufacturer (Section G)
NAKANISHI INC.
700 shimohinata
kanuma-shi, tochigi-ken 322-8 666
JA   322-8666
Manufacturer Contact
sean kaufman
1800 global parkway
hoffman estates, IL 60192
2245128921
MDR Report Key23034849
Report Number9611253-2025-00045
Device Sequence Number9329317
Product Code EGS
UDI-Device Identifier04560264566002
UDI-Public(01)04560264566002
Combination Product (Y/N)N
Initial Reporter StateNV
Initial Reporter CountryUS
PMA/510(K) Number
K182999
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative,Distributor
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 11/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberNANO95LS
Device Catalogue NumberC1099
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/19/2025
Supplement Date Received by Manufacturer10/20/2025
Initial Report FDA Received Date09/12/2025
Supplement Report FDA Received Date11/09/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/13/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age56 YR
Patient SexFemale
Patient EthnicityNon Hispanic
Patient RaceAsian
-
-