• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER LEIBINGER FREIBURG FACIAL ID-ORTHOG-6PLATES; IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER LEIBINGER FREIBURG FACIAL ID-ORTHOG-6PLATES; IMPLANT Back to Search Results
Catalog Number 78-92006
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Pain (1994)
Date of Event 08/12/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the facial id plates were explanted and found to be broken.
 
Additional Manufacturer Narrative
Device is not available for evaluation.If additional information is received, it will be reported on a supplemental report.
 
Additional Manufacturer Narrative
Additional information: h4, h6 device evaluation: follow-up report submitted to document device evaluation results.The reported event could be confirmed according to visual and dimensional inspection of the plate fragments returned.The investigation found that plate #1 initially fractured due to overload conditions, with sem analysis revealing low-cycle fatigue characteristics, while other plates showed forced fractures and severe embrittlement.Missing screws in upper holes and the use of an off-label screw with a larger diameter reduced the stability of the plate-screw-bone construct, likely causing micromotions and loosening of screws.Strong deformation of plates and damage to several screws occurred during removal, making it impossible to determine the exact location of the improper screw.Based on the investigation, there is no indication for an incorrectly working product or any design, material, or manufacturing-related issue.Therefore, no corrective and/or preventive actions are deemed necessary at this time.
 
Event or Problem Description
No new information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FACIAL ID-ORTHOG-6PLATES
Common Device Name
IMPLANT
Manufacturer (Section D)
STRYKER LEIBINGER FREIBURG
boetzingerstr. 41
freiburg D-791 11
GM  D-79111
Manufacturer (Section G)
STRYKER LEIBINGER FREIBURG
boetzingerstr. 41
freiburg D-791 11
GM   D-79111
Manufacturer Contact
elena drewenskus
boetzingerstr. 41
freiburg D-791-11
GM   D-79111
7614888208
MDR Report Key23034941
Report Number0008010177-2025-00108
Device Sequence Number13093650
Product Code JEY
UDI-Device Identifier07613327370867
UDI-Public07613327370867
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/11/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Catalogue Number78-92006
Device Lot Number2104261036
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 08/21/2025
Supplement Date Received by Manufacturer08/21/2025
Initial Report FDA Received Date09/12/2025
Supplement Report FDA Received Date12/11/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/14/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-