| Model Number |
8621500 |
| Medical Device Problem Codes |
Gas Output Problem (1266); Failure to Analyze Signal (1539); Failure to Deliver (2338)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/05/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that towards the end of closing wound, there was an unusual noise from the device and the mechanical ventilation stopped.Manual ventilation was initiated, and the operation was completed safely.No patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation was performed based on the given information (including electronic log file).The reported issue could be confirmed and it was determined that a failure of the optical incremental encoder was the root cause.The device detected the failure and reacted as specified for this situation.In case of such failure, atlan will still provide fresh gas (including agent if a vaporizer is connected) and monitoring allowing for manual ventilation.Since similar complaints have been reported recently a capa (corrective and preventive actions) was initiated.
|
| |
|