| Model Number |
MX40 1.4 GHZ SMART HOPPING |
| Medical Device Problem Code |
No Audible Alarm (1019)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/20/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Diagnostic/functional testing was performed at the philips authorized repair facility.Results of functional testing indicate that the speaker doesn't work in the unit.It states there is a speaker malfunction on the pic station and the speaker doesn't work with the fixture.The speaker was replaced.The device was operational after repairs were completed and the device was returned to the customer.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
|
| |
|
Event or Problem Description
|
|
Philips received a complaint on the mx40 1.4 ghz smart hopping device indicating that, during evaluation at bench repair, the device had no sound.The device was not in use on a patient at the time of the event, there was no patient involvement.
|
| |
|