The lens was returned positioned incorrectly in the lens case inside the lens carton.Viscoelastic was dried on the lens.One haptic was bent in the gusset areas with the distal area folded across the optic edge, similar to have been loaded into a cartridge.The lens was cleaned with klrp to further evaluate.The bent haptic relaxed into the original position after the removal of the dried viscoelastic.The other haptic was scraped across the anterior gusset area.A dimensional inspection was conducted.The lens dimensions (plan view) met specifications per an approved template.A device history record review and a non-conformance review of the reported lot number was conducted.The deviation review did not reveal any potential contributing factors to the reported complaint and all corresponding production release specifications defined in the device master record were met.Associated product information was not provided.It is unknown if qualified products were used.The root cause is most likely related to user handling.One haptic was bent and folded onto the optic edge, similar to have been loaded into a cartridge.The other haptic was scraped.Viscoelastic was dried on the lens.This was most likely interpreted as the reported complaint of "looked defective".All lenses are 100% inspected for cosmetic attributes and the damage exhibited by the returned complaint sample would not have met company's release criteria.Based on our observation, it can be reasonably concluded that the damage is not manufacturing related.Due to the presence of surgical solution and the condition of the returned sample, the damage is most likely related to customer handling.The lens was cleaned to remove the viscoelastic.The bent haptic relaxed into the original position.A dimensional inspection was conducted.The lens dimensions (plan view) met specifications per an approved template.It is unknown if qualified associated products were used.The instructions for use (ifu) instructs: company foldable intraocular lens (iols) are qualified for use with a company qualified delivery system (handpiece and cartridge) and ophthalmic viscosurgical device (ovd) combination.The ifu instructs that a company qualified delivery system and viscoelastic combination should be used.The use of an unqualified combination may cause damage to the lens and potential complications during the implantation process.The ifu instructs: using holding forceps, grasp the lens by the optic edge and gently place the lens anterior side up into the back of the ovd-filled cartridge.The lens should be inserted until the optic is a little more than half-way inside the cartridge.Use the holding forceps to gently push down on the lens, verifying that the lens is on the bottom surface of the cartridge.Using holding forceps, take the trailing haptic, and gently fold the haptic onto the anterior side of the optic.Slowly grip or push the optic edge to position the lens as far into the cartridge as the forceps will permit, while ensuring the lens remains on the bottom surface of the cartridge and the trailing haptic remains on the optic.Failure to follow these steps may cause the lens to advance incorrectly causing delivery issues and/or damage.The ifu instructs: follow the section regarding directions for use for information on the maximum allowed time for the iol to stay in the folded condition.Failure to adhere to manufacturer¿s recommendations may result in iol damage.Ifu note: during lens loading and insertion, do not allow the company iol to remain in a folded condition within the selected iol delivery system for more than 3 minutes prior to completing insertion into the capsular bag.The manufacturer internal reference number is: (b)(4).
|