TERUMO PHILIPPINES CORPORATION SURGUARD3 SAFETY HYPODERMIC NEEDLE; NEEDLE, HYPODERMIC, SINGLE LUMEN Back to Search Results
Model Number SG3-2525
Medical Device Problem Code Malposition of Device (2616)
Health Effect - Clinical Code Needle Stick/Puncture (2462)
Date of Event 08/14/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
E3: occupation: administrative assistant osha safety officer the details of the actual condition were unable to be determined as the actual sample was not available for evaluation.Retention samples were visually inspected and found to be free of any damage or irregularities on the sheath and collar.There was no issued nonconformity report related to human error during lot production.The needles were manufactured at needle assembly machine 8, which has an inspection system that detects tilted and bent cannulas.Dummy checking is conducted for every type of change, end of the lot, and machine adjustment/troubleshooting/replacement of parts to ensure the accuracy of the inspection system.During the collar assembly process, the hub fit is loosened from the protector before placing the needle into the collar.Before proceeding to the next process, the hub, cannula, and collar are pressed into the protector, wherein a photoelectric sensor detects if the height of the protector is correct.Our safety needle features a hinged safety sheath attached to the needle hub.The safety sheath contains two locking mechanisms, the sheath tooth-cannula and sheath wings collar, which are simultaneously activated when manually pressed over the needle immediately after use and just before disposal to minimize the possibility of sharps injury.Our product has an allowable tilting of the needle of not more than 2°.Tilted cannulas are one of the check items in the hourly in-process inspection during needle assembly.The collar and sheath undergo an initial product inspection check (ipic) before mass production, during raw material change, mold maintenance, plant shutdown, cavity closure, and mold/product type change.Any molding defects that may affect product function are part of the inspection item.We also have a visual in-process and product confirmation inspection to ensure that the collar and sheath are in good condition.Visual in-process inspections are conducted during the manufacturing process to detect abnormalities on the product that may cause issues during sheath activation.The critical locking part of the sg3 product is checked for defects such as short shot, sheath collar fitting, and over- or under-insertion of the collar into the hub.We have functional tests conducted to confirm that our products are free of defects that would lead to difficulty or failure of device activation.We perform safety sheath activation during sg assembly inspection and outgoing inspection of finished products, where the cannula should be captured under the sheath's tooth.During the component fit test, we also check that the sheath lugs are securely attached to the collar ear.Before shipment, qc conducts outgoing inspections to check the overall condition of the products.The product's instructions for use (ifu) provide detailed instructions for using the sg3 safety needle device, including warnings to prevent needle sticks.Our cannula is a supplied raw material that complies with iso 9626, particularly on stiffness and breakage.It undergoes incoming inspection, which includes visual inspection, dimensional inspection, and verification of the quality certificate.From the previous two fiscal years to the present, we received a total of seventy-one (71) complaints for the same issue affecting 25g needles.The cause of these complaints was not identified as being related to our product or the manufacturing process.The retention samples were injected into the dummy arm's veins and a rubber cork to simulate intravenous and intramuscular injections prior to safety activation.The safety sheath was successfully activated on the samples; the needle did not bend, and no difficulties were encountered.In the next simulation, the needle was forcefully injected into the cork, causing it to slightly bend.Despite being bent, the sheath-tooth was able to completely capture the cannula, and the collar wing was fully locked on the sheath, indicating that the activation was successful.We could not identify the root cause of the problem based on our investigation of all the available information regarding this issue.A review of the product's lot history file revealed that no related issues were encountered during the production of the reported lot.We have routine inspections to check the condition of the products.The components that are critical to the product's safety activation are checked to ensure that no defects occur that could result in sheath activation issues.We recommend following the instructions for use (ifu) for proper use of the sg3 needle, which includes caution, precautions, and warnings to avoid problems during sheath activation.Terumo medical corporation (tmc) (importer) registration no.2243441 is submitting this report on behalf of terumo (philippines) corporation (manufacturer) registration no.3003902955.
 
Event or Problem Description
Terumo medical received reported information indicating that the safety mechanism failed to activate properly after use, resulting in the needle bending and remaining exposed.Additional information received on (b)(6) 2025 stated that the procedure involved vaccine administration to a pediatric patient.During activation of the safety sheath against a hard exam tray, the needle bent and was only partially captured inside the sheath, leaving the tip exposed.A needle stick injury occurred, resulting in staff exposure to the needle tip.The affected staff member required laboratory testing for potential exposure to bloodborne pathogens and was able to return to normal duties following the incident.The pediatric patient was not injured.The concomitant device used was a prefilled syringe containing the vaccine components.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURGUARD3 SAFETY HYPODERMIC NEEDLE
Common Device Name
NEEDLE, HYPODERMIC, SINGLE LUMEN
Manufacturer (Section D)
TERUMO PHILIPPINES CORPORATION
124 east main avenue
laguna technopark
binan laguna, 4024
RP  4024
Manufacturer (Section G)
TERUMO PHILIPPINES CORPORATION
124 east main avenue
laguna technopark
binan laguna, 4024
RP   4024
Manufacturer Contact
stephanie handy
950 elkton blvd.
elkton, MD 21921
9499890491
MDR Report Key23035815
Report Number3003902955-2025-00066
Device Sequence Number21210992
Product Code FMI
UDI-Device Identifier34806017508204
UDI-Public(01)34806017508204(17)280630(10)230712E
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
K122249
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/12/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSG3-2525
Device Catalogue NumberN/A
Device Lot Number230712E
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/14/2025
Initial Report FDA Received Date09/12/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/12/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PREFILLED SYRINGE WITH VACCINE COMPONENTS