Model Number 1020007
Medical Device Problem Codes Overheating of Device (1437); Failure to Deliver (2338)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/04/2025
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer received information alleging no oxyge on a everflo device.There was no reported harm.The device was returned to the manufacturer's service center for evaluation, and the customer¿s complaint was confirmed.During the evaluation, it was noted that the device had a burned pca and the device does not power on.The root cause is confirmed to be thermal damage of the pca.Additionally, during the service of the device, the manufacturer observed a dust-like contamination on the rear enclosure, scratches on the front enclosure, brown discoloration on the compressor inlet tube (indicative of tobacco use).Kinked/pinched cannula tube.Thermal damage on the pca (f3, t3) and blown fuse (f1) which is indicative of a voltage spike that occurred external to the device.Pil was unable to power on and test the device due to thermal damage on the pca.At this time, no further investigation can be performed.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVERFLO
Common Device Name
GENERATOR, OXYGEN, PORTABLE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
7247330200
MDR Report Key23036716
Report Number2518422-2025-110091
Device Sequence Number21210912
Product Code CAW
UDI-Device Identifier00606959032569
UDI-Public(01)00606959032569
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K061261
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation 505
Type of Report Initial
Report Date (Section B) 09/12/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number1020007
Device Catalogue Number1020007
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/21/2025
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/04/2025
Initial Report FDA Received Date09/12/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/30/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-