| Catalog Number |
89-8509-450-60 |
| Medical Device Problem Code |
Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/18/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).G2 ¿ foreign ¿ ecuador investigation of this incident is currently ongoing.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that during regular inspection of warehouse equipment, it was identified that the oscillating saw attachment was working normally until the first test cut, but by the time the second cut is made, the attachement stopped working.The trigger was engaged, it was turning but had no power, and it sounded as if there's something loose inside the motor that's preventing any quick-release coupling or saw attachment from turning.No consequences or impact to the patient.Due diligence is in progress for this complaint; to date no additional information or product has been received.
|
| |
|
Additional Manufacturer Narrative
|
|
The following sections were updated: b4, b5, d2, d4, g3, g6, h2, h3, h4, h6, h11.The following sections were corrected: h6 component codes.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history record identified no deviations or anomalies during manufacturing.Review of complaint history identified no additional similar complaints for the reported item and no additional complaints for the reported part and serial combination.Complaints are monitored through monthly complaint review in order to identify potential adverse trends.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Search Alerts/Recalls
|