| Model Number |
KD-VC611Q-07303S |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/22/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported the single use preloaded sphincterotome cutting wire snapped.The issue occurred during a therapeutic endoscopic retrograde cholangiopancreatography (ercp)procedure.The procedure was completed using a similar device.There were no reports of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
The mdr supplemental report is being submitted to provide updated fields and final investigation results.Updated fields: d8, d9, h2, h3, h4, h6, h11.The device was returned for evaluation and the customer's allegation was confirmed.A definitive root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: the device was not protruded enough from the endoscope until the rear end of the cutting wire was in the field of view.The cutting wire and the endoscope were being close to each other.The output was activated.This had led to an electrical discharge between the cutting wire and the distal end of the endoscope.An electrical discharge occurred, and the cutting wire became hot instantly.This had caused the cutting wire to break.Olympus will continue to monitor the performance of this device.
|
| |
|
Event or Problem Description
|
|
No additional information received from the customer.
|
| |
|
Search Alerts/Recalls
|