|
Additional information provided in d.9., h.3., h.6.And h.11.The used pak was visually inspected and no obvious defects were found.The ball in the drip chamber¿s check valve moved freely per specification.The small parts tray was returned.The returned manifolds, connectors, and components were found to be in good condition.The connectors were connected to the cassette and handpiece without any interference.The returned product was tested using a calibrated console.The product could prime and tune with the ultrasonic handpiece, the 0.9 milli meter aspiration bypass system tip and returned infusion sleeve successfully.Fluid flowed from the balanced salt solution bottle to the drain bag without any interfering.No message code appeared on the screen.No bubble or air leakage was observed from the returned cassette, manifolds, connectors, and components.No leakage was detected from the pump elastomer or on the pump area of the fluidics module.The product passed functionality testing.The investigation conducted a non-conformance review of the reported lot number.No deviations that could have contributed to this event were identified and all corresponding production release specifications defined in the device master record were met.Based on the evaluation of the information and materials received, the investigation was unable to identify the root cause or origin of the reported event.The product functioned per specifications.Additionally, no manufacturing-related deficiencies were found that potentially could have contributed to the complaint.No further investigative or manufacturing actions are warranted at this time due to the product meeting specifications.Receipt of additional relevant information or supporting materials will prompt a re-evaluation of the complaint investigation.Complaint data for all company products is reviewed monthly to monitor for adverse trends.During the last review, no adverse trends were observed for the reported product and event combination.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
|