Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: h3, h6 investigation summary the product was not returned to j&j medtech orthopaedics, however photos were provided for review.The photo investigation revealed investigation revealed nothing indicative of a device nonconformance; device was observed implanted and explanted.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.The overall complaint was not confirmed as the observed condition of the rfna/ 14mm/ 380mm/ standard bend/ ster would have not contributed to the reported adverse event.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review (dhr): manufacturing location: monument.Manufacturing date: 22-apr-2023.Expiration date: 31-mar-2028.Part number: 04.233.438s, rfn/14mm/380mm std bend/pe inlay/sterile.Lot number: 5767p51 (sterile) lot quantity: 12.Nonconformance noted: n/a.A manufacturing record evaluation was performed for the finished device with batch number 5767p51.No nonconformities or manufacturing irregularities have identified related to the complaint condition.
|