MAUDE Adverse Event Report: SYNTHES GMBH RFNA / 14MM / 380MM STANDARD BEND / STERILE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH RFNA / 14MM / 380MM STANDARD BEND / STERILE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Catalog Number 04.233.438S
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Nonunion/Delayed-union Bone Fracture (2369)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: h3, h6 investigation summary the product was not returned to j&j medtech orthopaedics, however photos were provided for review.The photo investigation revealed investigation revealed nothing indicative of a device nonconformance; device was observed implanted and explanted.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.The overall complaint was not confirmed as the observed condition of the rfna/ 14mm/ 380mm/ standard bend/ ster would have not contributed to the reported adverse event.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review (dhr): manufacturing location: monument.Manufacturing date: 22-apr-2023.Expiration date: 31-mar-2028.Part number: 04.233.438s, rfn/14mm/380mm std bend/pe inlay/sterile.Lot number: 5767p51 (sterile) lot quantity: 12.Nonconformance noted: n/a.A manufacturing record evaluation was performed for the finished device with batch number 5767p51.No nonconformities or manufacturing irregularities have identified related to the complaint condition.
 
Event or Problem Description
It was reported that the patient had hardware removal of rfna and was revised due to non-union.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RFNA / 14MM / 380MM STANDARD BEND / STERILE
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SYNTHES GMBH
luzernstrasse 19-21
zuchwil 4528
SZ  4528
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
elba bello
1302 wright lane east
west chester, PA 19380
9083863534
MDR Report Key23038345
Report Number8030965-2025-09264
Device Sequence Number9476625
Product Code HSB
UDI-Device Identifier10886982291165
UDI-Public(01)10886982291165(17)280331(10)5767P51
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
K233696
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number04.233.438S
Device Lot Number5767P51
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/27/2025
Initial Report FDA Received Date09/12/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/22/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-