| Catalog Number |
865034 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/13/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: d4: udi: as a valid lot number for the device involved in the event was not provided, the full udi is currently not available.H4: manufacture date is an unknown day in august 2009.
|
| |
|
Event or Problem Description
|
|
It was reported that the arm on patella guide dislodged after cutting the patella from the vibration of the saw.No injury or delay happened.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary- it was reported that the arm on patella guide dislodged after cutting the patella from the vibration of the saw.No injury or delay happened.The product was returned to j&j medtech orthopaedics for evaluation.Visual analysis of the returned device found that the pfc*calibrated pat cut gde has one of the two saw captures broken, the broken fragment was returned for inspection.Additionally the device exhibits heavy signs of use for a long period of time.The observed condition was identified as an end of life indicator; damage consistent with repeated use and servicing.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use, the device must be properly inspected prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions, and inspection procedures.A functional test was not performed since it was not applicable to the complaint condition.A dimensional inspection was not performed since it was not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the pfc*calibrated pat cut gde would have contributed to the complained issue.Based on the investigation findings, the potential cause is traced to end of life, and it has been determined that no corrective and/or preventative action is required.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot- a device history record (dhr) review or manufacturing records evaluation (mre) was not possible because the date code j0809 provided is not a valid finished goods lot number.H11 additional narrative: corrected: h3.
|
| |
|
Search Alerts/Recalls
|