• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JJGC S.A. GM ZYGOMATIC IMPLANT, TI, 4.0X30; ENDOSSEOUS DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JJGC S.A. GM ZYGOMATIC IMPLANT, TI, 4.0X30; ENDOSSEOUS DENTAL IMPLANT Back to Search Results
Catalog Number 109.1049
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/09/2025
Type of Reportable Event Serious Injury
Event or Problem Description
The clinician reports that the day the implant was placed in ada 3, failure occurred upon insertion.Details of surgery: primary stability not achieved.Patient presented with inadequate bone quality/quantity.The device was forwarded to the manufacturer.There were no reported patient injuries or complications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GM ZYGOMATIC IMPLANT, TI, 4.0X30
Common Device Name
ENDOSSEOUS DENTAL IMPLANT
Manufacturer (Section D)
JJGC S.A.
av. juscelino kubitschek de
oliveira, 3291 cic
curitiba PR 81270 -200
BR  81270-200
MDR Report Key23040667
Report Number0001222315-2025-002578
Device Sequence Number10096132
Product Code DZE
UDI-Device Identifier07899878031888
UDI-Public07899878031888
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Importer
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 09/12/2025
Report Date (Section F) 08/19/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number109.1049
Device Lot NumberTGT88
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/19/2025
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date09/12/2025
Device Age30 MO
Event Location Other
Date Report to Manufacturer08/19/2025
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date09/12/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-