JJGC S.A. CM ALVIM ACQUA IMPL TI 4.3X16; ENDOSSEOUS DENTAL IMPLANT Back to Search Results
Catalog Number 140.651
Medical Device Problem Code Failure to Osseointegrate (1863)
Health Effect - Clinical Code Pain (1994)
Date of Event 07/10/2025
Type of Reportable Event Serious Injury
Event or Problem Description
The clinician reports the implant was inserted on (b)(6) 2025 in the patient's mouth.Details of surgery: immediate extraction site.On (b)(6) 2025, non-osseointegration was verified.Patient presented with inadequate bone quality/quantity.The device was forwarded to the manufacturer.At the event the patient experienced: pain.No further patient complications were reported.
 
Additional Manufacturer Narrative
The batch number could be verified.Our manufacturing q-system assures, that production and process controls are in place to ensure that batches confirm to the applicable specifications before they are distributed.The non-integration of an endosseous dental implant during the healing phase is a known inherent risk of the treatment with dental implants.Most implant failures occur before occlusal loading (analainen et al.2009).Based on clinical studies about 1-3% of the implants fail within the first year after implantation (ganeles et al.2008).Implants may have to be removed in case one or more of the implant success criteria are not met.Implant success criteria according to buser et al.(1991) are: 1.Absence of persistent subjective complaints such as pain, foreign body sensation and /or dysesthesia.2.Absence of a recurrent peri-implant infection with suppuration.3.Absence of implant mobility.4.Absence of a continuous radiolucency around the implant.The manufacturer's trend analysis confirms that the reported early failure rate associated with its dental implants is below the expected failure rate for this treatment as published in the scientific literature.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CM ALVIM ACQUA IMPL TI 4.3X16
Common Device Name
ENDOSSEOUS DENTAL IMPLANT
Manufacturer (Section D)
JJGC S.A.
av. juscelino kubitschek de
oliveira, 3291 cic
curitiba PR 81270 -200
BR  81270-200
Manufacturer (Section G)
JJGC S.A.
av. juscelino kubitschek de
oliveira, 3291 cic
curitiba PR 81270 -200
BR   81270-200
Manufacturer Contact
jennifer jackson
60 minuteman road
andover, MA 01810
9787472509
MDR Report Key23041505
Report Number3008261720-2025-003293
Device Sequence Number9889467
Product Code DZE
UDI-Device Identifier07899878016632
UDI-Public07899878016632
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K150199
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 09/12/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number140.651
Device Lot NumberJXPX6
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/18/2025
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date09/12/2025
Event Location Other
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date09/12/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/05/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age52 YR
Patient SexMale
-
-