| Model Number |
H74939186401010 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/20/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
B3 - date of event: date of event was approximated to 08/01/2025 as the exact event date was not reported.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.4.0mm x 100mm x 150cm coyote balloon catheter was advanced for dilation.During procedure the balloon burst during inflation without reaching the burst rated pressure.The procedure was completed.There were no patient complications as a result of this event.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.4.0mm x 100mm x 150cm coyote balloon catheter was advanced for dilation.During procedure the balloon burst during inflation without reaching the burst rated pressure.The procedure was completed.There were no patient complications as a result of this event.It was further reported that the patient was treated for a peripheral arterial disease at the superficial femoral artery (sfa) for the sterling balloon and tibial artery for the coyote.The target lesion was approximately above 90% stenosis, not tortuous, and moderately to severely calcified.The balloon ruptured at the first inflation while taking the it up at below burst pressure.
|
| |
|
Additional Manufacturer Narrative
|
|
B3 - date of event: date of event updated.
|
| |
|