BIOSENSE WEBSTER INC PENTARAY NAV; CATHETER, INTRACARDIAC MAPPING, HIGH-DENSITY ARRAY Back to Search Results
Catalog Number D128208
Medical Device Problem Code Obstruction of Flow (2423)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/26/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that it was noticed that the pentaray nav was not flushing at all before readvancing the pentaray nav back into the body for a post-map.The pentaray nav was flushed but the issue persisted.The pentaray nav was replaced and the procedure continued with no further issues.The caller stated that the faulty pentaray nav was brand new.There was no patient consequence.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster inc., or its employees that the report constitutes an admission that the product, biosense webster inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's ref.# (b)(4).
 
Additional Manufacturer Narrative
It was reported that it was noticed that the pentaray nav was not flushing at all before readvancing the pentaray nav back into the body for a post-map.Device investigation details: an analysis of the product could not be performed since a physical sample was not received for evaluation.However, if the product is received at a later date, the investigation will be updated as applicable.A manufacturing record evaluation was performed for the finished device number lot 31700005l and no internal action related to the complaint was found during the review.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTARAY NAV
Common Device Name
CATHETER, INTRACARDIAC MAPPING, HIGH-DENSITY ARRAY
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
henry boden
31 technology dr
irvine, CA 92618
9497882078
MDR Report Key23042864
Report Number2029046-2025-03133
Device Sequence Number9870924
Product Code MTD
UDI-Device Identifier10846835012224
UDI-Public(01)10846835012224(17)280707(10)31700005L
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K123837
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberD128208
Device Lot Number31700005L
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/26/2025
Supplement Date Received by Manufacturer11/05/2025
Initial Report FDA Received Date09/12/2025
Supplement Report FDA Received Date11/19/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/08/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MEDTRONIC PFA SYSTEM.
Patient SexUnknown
-
-