| Catalog Number |
SRM-640 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
07/22/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Event or Problem Description
|
|
Patient reported "rupture" diagnosed via mri.Healthcare professional confirmed rupture.This record is for the left side.Device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the product analysis performed, the assessments of the complaints are: ¿ rupture: observed broken device assessed as fold flaw opening.As per the investigation procedure, stress marks was observed and none of the observations are found to be potentially related to the manufacturing process, no further actions are required.
|
| |
|
Event or Problem Description
|
|
Patient reported "rupture".Diagnosed via mri.This record is for the left side.Device was explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: b5, b6, d6b, h3, h6.
|
| |
|
Event or Problem Description
|
|
Patient reported "rupture" diagnosed via mri.Healthcare professional confirmed rupture.This record is for the left side.Device was explanted and replaced.
|
| |
|