| Model Number |
EGIA30CTAVM |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Hemorrhage/Blood Loss/Bleeding (1888); Laceration(s) (1946); Tissue Breakdown (2681)
|
| Date of Event |
08/21/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that during a video-assisted thoracoscopic surgery (vats) right lower lobe lobectomy, while performing ligation of the pulmonary artery, immediate bleeding occurred following stapler firing.The procedure was converted from laparoscopic to open thoracotomy to achieve hemostasis.The estimated blood loss was approximately 1500 milliliters (ml).Upon inspection after vascular suturing, the staple line at the distal pulmonary artery (specimen side) was complete; however, no staple line was observed on the proximal pulmonary artery (patient side), and a sharp transection was identified.It was noted that the staple line did not hold or opened up, and suturing was required as a replacement for the staple line.The patient remained hospitalized.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d4(udi), d9, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the staple line did not hold.The reported issue could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|