• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIODYNAMICS, INC SMART PORT CT; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIODYNAMICS, INC SMART PORT CT; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR Back to Search Results
Model Number H787CT75STSDVI1
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Medical device site infection (4845)
Date of Event 09/30/2015
Type of Reportable Event Serious Injury
Event or Problem Description
On (b)(6) 2015, (b)(6).Underwent placement of an angiodynamics' smartport ct device, catalog # ct75stsd-vi, lot # 4792614.Upon information and belief, the device's description is as follows: smartport ct single titanium port system with attachable 7.5f x 66cm silicone catheter and 8f introducer kit (with valved introducer).The device at issue was implanted by dr.(b)(6), m.D., (b)(6) hospital, in (b)(6), for the treatment of plaintiff's medical condition.On (b)(6) 2015, (b)(6).Presented to (b)(6) hospital in (b)(6).At that time, (b)(6) medical team determined that her smartport ct was infected and that the infected port-a-cath had to be removed.On (b)(6) 2015, (b)(6) defective port was removed by dr.(b)(6), md, at (b)(6) hospital, in (b)(6).On (b)(6) 2015, the post-operative diagnosis revealed that (b)(6).'s port-a-cath at issue was infected.After removal of the infected smartport ct port-a-cath, plaintiff did not have an opportunity to consult with dr.(b)(6), m.D.About the causes of her catheter infection and her post-operative diagnosis of the infected port-a-cath.Plaintiff knew she had an infection, but she did not know her infection was caused by the smartport ct's defects and/or by the defendants' conduct.Plaintiff did not discover the heightened risk of infections related to the smartport ct's defects until around (b)(6) 2023, when plaintiff saw a social media ad which detailed potential defects in port catheter products.
 
Additional Manufacturer Narrative
The customer's reported complaint description of patient infection cannot be confirmed given the patient centric nature of this serious adverse event (sae).No port device was returned for evaluation since there was no report of port device malfunction, damage or performance issue during in situ use/treatment.A device history record (dhr) review of the indicated packaging lot revealed no quality related issues or manufacturing deficiencies at the time of manufacture.Certificate of conformance for the sterilization process for the reported packaging lot was verified to be approved at time of distribution.The port was implanted into the patient more than 6 months prior to the patient's infection.There was no report of port device malfunction or performance issue during use.If the manufacturing/packaging/sterilization of the port device in question was to contribute to an infection in the patient it would have likely occurred shortly after the implant date and not more than 6 months later.Device directions for use (dfu) cautions that each access of the port should be performed using aseptic technique, following the institutions universal precautions.There is no indication from the reported complaint that the manufacturing/packaging/sterilization of the port device could have contributed to the patient's infection.Infection is cautioned in the device dfu as a potential complication of port system use.Labeling review: the directions for use (dfu) that is provided with the port device contains the following statements: warnings: the device is to be implanted, used, maintained, and removed in strict accordance with institutional and or centers for disease control (cdc) guidelines or policies.Reuse of single-use devices creates a potential risk of patient or user infections.Contamination of the device may lead to injury, illness or death of the patient.Do not use syringes smaller than 10 ml syringe when accessing the port as system damage can occur.Flushing occluded catheters with small syringes can create excessive pressures within the port system.Precautions carefully read and follow all instructions prior to use.Strict aseptic technique is of paramount importance when implanting any device.For peripheral placement, irritation to the vein, resulting in postoperative thrombophlebitis, has been associated with guidewire and introducer insertion.When using percutaneous introducers: to avoid blood vessel damage, do not allow the percutaneous introducer sheath to remain indwelling in the blood vessel without the internal support of a catheter or dilator.If more than one drug is to be administered, between the individual drug applications, flush the system with 5 to 10 ml normal saline for injection to prevent drug interactions.After any infusion, injection or bolus application, the system should be flushed with normal saline for injection or locked with a heparin solution per institutional protocol to prevent thrombotic occlusion of the catheter.Potential complications use of an angiodynamics port system involves potential risks normally associated with the insertion or use of any implanted device or indwelling catheter including but not limited to: air embolism, catheter disconnection or migration, catheter embolization, catheter fragmentation, catheter pinch-off, clot formation, drug extravasation (leakage), erosion of vessel and skin, implant rejection, infection, inflammation, thromboembolism, thrombophlebitis, thrombosis, necrosis of scarring of skin over implant area, vessel trauma.Post-operative care: the incision site should be monitored for signs of infection, inflammation, hematoma, device rotation or erosion.Routine wound care should be given to these sites.The smart port ct implantable port may be used immediately after verification of catheter placement.Instruct patient to avoid any heavy exertion or strenuous activity during the first few days after surgery.General guidelines each access of an angiodynamics smart port ct implantable port should be performed using aseptic technique.Follow institutional universal precautions.A review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint continues to be monitored for trends.Reference (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMART PORT CT
Common Device Name
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
ANGIODYNAMICS, INC
10 glens falls technical park
glens falls NY 12801
Manufacturer (Section G)
ANGIODYNAMICS, INC
10 glens falls technical park
glens falls NY 12801
Manufacturer Contact
alexandra invencio
26 forest street
marlborough, MA 01752
5086587990
MDR Report Key23048097
Report Number1317056-2025-00228
Device Sequence Number15961979
Product Code LJT
UDI-Device Identifier15051684018135
UDI-Public15051684018135
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K062414
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2015
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 09/14/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/01/2017
Device Model NumberH787CT75STSDVI1
Device Catalogue NumberCT75STSD-VI
Device Lot Number4792614
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/20/2025
Initial Report FDA Received Date09/14/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/10/2014
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Other; Hospitalization;
Patient SexFemale
-
-