| Model Number |
45-781639-L |
| Medical Device Problem Code |
Premature Separation (4045)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/22/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
There are three coils associated with the reported event.This report addresses the third coil.The first coil is referenced under mfr report # 2032493-2025-90481.The second coil is referenced under mfr report # 2032493-2025-90482.The device was discarded at the user facility and not returned to the manufacturer for evaluation.Procedural or medical imaging were requested and not provided.Without the return and physical evaluation of the device, the investigation is unable to determine if a condition existed that would have caused or contributed to the reported event.The event as described could not be confirmed.If imaging is received at a later date, an investigation will be performed and a supplemental mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that upon gaining access to the aneurysm, the coil was prepped and introduced into the microcatheter and when it encountered the gentle curves of the splenic artery it began to bind.The coil was removed and the microcatheter was flushed.The physician was able to remove 1/3 of the coil out, but upon attempt to position the coil, it became detached from the pusher wire.The entire micro system was removed.It was reported that this issue occurred with three separate coils during the case.The physician switched to a competitor's coils and was able to successfully treat the aneurysm.There was no patient injury reported.The patient recovered well without issues.
|
| |
|
Search Alerts/Recalls
|