MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HOUSTON CONSTELLATION SURGICAL PROCEDURE PACK; GENERAL SURGERY TRAY (KIT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HOUSTON CONSTELLATION SURGICAL PROCEDURE PACK; GENERAL SURGERY TRAY (KIT) Back to Search Results
Catalog Number 8065751155
Medical Device Problem Code Obstruction of Flow (2423)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/30/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A physician reported that the ultrasound tip could not be oscillated during the surgery.The surgery was completed on same day by replacing the product.The procedure type was unknown.There was no patient impact.
 
Additional Manufacturer Narrative
No technical inquiries were received from the customer.A sample was not received at the manufacturing site for evaluation; therefore, the condition of the product could not be verified.The reported product¿s lot number did not contain a phacoemulsification (phaco) tip, preventing lot number specific reviews for similar complaints or nonconformances.However, before production release, each product history record is reviewed to ensure that all associated products meet the required specifications and release criteria.Based on the evaluation of the information received, the investigation was unable to identify the root cause or origin of the reported event.As the root cause and its origin are inconclusive, further investigative or manufacturing actions are not warranted at this time.All phaco tips are 100% visually inspected by trained operators using 30x magnification during the manufacturing process.Receipt of additional relevant information or supporting materials will prompt a re-evaluation of the complaint investigation.Complaint data for all products is reviewed monthly to monitor for adverse trends.During the last review, no adverse trends were observed for the reported product and event combination.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION SURGICAL PROCEDURE PACK
Common Device Name
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key23048900
Report Number1644019-2025-03398
Device Sequence Number9741426
Product Code LRO
UDI-Device Identifier00380657511556
UDI-Public(01)00380657511556(17)261231(11)250104(10)176UCM
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K880961
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/04/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number8065751155
Device Lot Number176UCM
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 08/20/2025
Supplement Date Received by Manufacturer11/22/2025
Initial Report FDA Received Date09/15/2025
Supplement Report FDA Received Date12/04/2025
Date Device Manufactured01/04/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CONSTELLATION VISION SYSTEM.; HAND PIECE.
Patient SexUnknown
-
-