• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARDINAL HEALTH 200, LLC KANGAROO; PUMP, INFUSION, ENTERAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDINAL HEALTH 200, LLC KANGAROO; PUMP, INFUSION, ENTERAL Back to Search Results
Model Number 773656
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/12/2025
Type of Reportable Event Malfunction
Event or Problem Description
Feed pump alarming for feed error despite no apparent kinks or clogs.Feed bags changed multiple times ac well as feeding pump and problem continued.Line was flushed and primed to clear any potential clogs as well.The pump continued to alarm hourly until changed to a feed bag that included a flush bag.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KANGAROO
Common Device Name
PUMP, INFUSION, ENTERAL
Manufacturer (Section D)
CARDINAL HEALTH 200, LLC
3651 birchwood drive
waukegan IL 60085
MDR Report Key23050591
Report Number23050591
Device Sequence Number15894982
Product Code LZH
UDI-Device Identifier10192253009941
UDI-Public(01)10192253009941
Combination Product (Y/N)N
Initial Reporter StateVA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type User Facility
Initial Reporter Occupation Risk Manager
Type of Report Initial
Report Date (Section B) 08/15/2025
Report Date (Section F) 08/15/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number773656
Device Catalogue Number773656
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Event Location Hospital
Date Report to Manufacturer09/15/2025
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date09/15/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
Date Report Sent to FDA08/15/2025
-
-