|
|
| Catalog Number |
TSVTWH11 |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/05/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zimvie complaint number (b)(4).A4: weight unknown / not provided.G4: additional pma/510(k) number ¿ k133339.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the threads inside of the implant at tooth site #12 are stripped, not allowing the abutment to seat.Looking to get something to rethread the implant allowing for the abutment to seat.
|
| |
|
Event or Problem Description
|
|
No additional event information received at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted to relay additional information and device evaluation.The following sections are being reported: b4: date of this report was updated.G3: date received by manufacturer was updated.G6: type of report was updated.H2: type of follow up was updated.H3: device evaluated by manufacturer was updated.H6: type of investigation codes were added: 4109, 4111 and 3331.H6: investigation findings code was added: 213.H6: investigation conclusions code was added: 4315.H10: narrative/data was updated.Zimvie did not receive the reported implant for evaluation.Visual inspection could not be performed.The investigation has been performed based on the available information using the item number, device history record (dhr) review, and risk management file (rmf).Functional testing to recreate the reported event could not be performed due to the nature of the device & event.Dhr review was completed for the subject lot number.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number for similar events and no other complaint was identified.Based on the investigation and risk management file review, the most likely root causes determined from the investigation are incorrect techniques used - customer error.Therefore, based on the available information, device malfunction could not be verified.Without device receipt, the reported event is non-verifiable.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the product was nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.
|
| |
|
Search Alerts/Recalls
|
|
|