• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COLOPLAST A/S TITAN INFLATABLE PENILE PROSTHESIS FAMILY; PENILE PROSTHESIS, INFLATABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COLOPLAST A/S TITAN INFLATABLE PENILE PROSTHESIS FAMILY; PENILE PROSTHESIS, INFLATABLE Back to Search Results
Catalog Number ES8822
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/27/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The lot number was reviewed for complaint trend, non-conforming report and capa.Devices met specification prior to release.No trends were noted for complaints and there were no nonconforming reports or capas that were confirmed in veeva to be associated.
 
Event or Problem Description
According to the available information, the device was explanted and replaced due to a tubing break at the pump.
 
Additional Manufacturer Narrative
Reservoir was received for evaluation.Abrasion was noted on the inlet tubing of the reservoir.A separation was noted on the inlet tubing of the reservoir at the strain relief junction.This was a site of leakage.The surfaces appear to be rough and irregular, indicating stress was exerted based on examination of the returned product, it was concluded that the rough and irregular surfaces associated with the separation indicates that sufficient stress was exerted on the inlet tube of the reservoir to separate the site while in-vivo.A separation of this type could then allow the loss of fluid, making the device inoperable.A review of the device history record confirmed the devices from this lot met all specifications prior to release.A review of the complaint history database, nonconformance's and capas revealed no trends for this lot.
 
Event or Problem Description
According to the available information, the device was explanted and replaced due to a tubing break at the pump.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TITAN INFLATABLE PENILE PROSTHESIS FAMILY
Common Device Name
PENILE PROSTHESIS, INFLATABLE
Manufacturer (Section D)
COLOPLAST A/S
1 holtedam humlebaek, dk 3050
humlebaek 3050
DA  3050
Manufacturer Contact
usbes brian schmidt
1601 west river road n
minneapolis, MN 55411
MDR Report Key23052087
Report Number2125050-2025-01463
Device Sequence Number9741334
Product Code FHW
UDI-Device Identifier05708932487269
UDI-Public05708932487269
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P000006
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/19/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue NumberES8822
Device Lot Number9286793_ES88222400
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/05/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date09/15/2025
Supplement Report FDA Received Date11/19/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/18/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age69 YR
Patient SexMale
-
-