| Catalog Number |
03.110.007 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/28/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported on (b)(6) 2025, the forcep's screw fell out, causing the device to split in two pieces.Also, silver cap on back of the screwdriver fell out.No operation involved.No consequences.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: added: d9.H3, h6: the device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: h3, h4, h6: the product was returned to j&j medtech orthopaedics for evaluation.The j&j medtech orthopaedics team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the cap of the sddrive screwdriver t8 was observed with a gap from the handle.A dimensional inspection for the sddrive screwdriver t8 was not performed due to post manufacturing damage.A functional test was performed and the cap is loose, movement was noted and could not be places correctly, however is not broken or fell apart.The complete potential cause cannot be established with the provided information.The overall complaint was confirmed as the observed condition of the sddrive screwdriver t8 would contribute to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.There is no indication that a design or manufacturing issue has caused the complaint condition.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review (dhr): part # 03.110.007, synthes lot # 6566935, suppliers lot # na, release to warehouse date: 19 jan 2011, manufactured by: synthes jennersville.No ncrs were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of this product, and any sub-components, which would contribute to this complaint condition.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|