• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CORDIS CORPORATION SABER035; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CORDIS CORPORATION SABER035; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Catalog Number 48007004R
Medical Device Problem Code Burst Container or Vessel (1074)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/21/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information is pending and will be submitted within 30 days upon receipt.
 
Event or Problem Description
As reported, a 7x40.035 saber percutaneous transluminal angioplasty (pta) balloon ruptured at nominal pressure after being inflated inside of a stent.The balloon ruptured below the rated burst pressure (rbp).The procedure was completed with post-dilatation using a saber.035 pta balloon.No harm or injury occurred to the patient.The product was stored properly per the instructions for use (ifu).The balloon was removed from the hoop, protective cover, and sterile packaging without difficulty.There were no kinks or damages were noted prior to insertion.The device was prepped according to the ifu and maintained negative pressure normally.The intended procedure targeted a lesion in the superficial femoral artery (sfa), which was severely calcified with no vessel tortuosity and 70¿80% stenosis.The device was not used for a chronic total occlusion (cto).No resistance or friction occurred during insertion through the rotating hemostatic valve or while advancing through the vessel or crossing the lesion.The catheter was not in an acute bend and did not kink during use.A 50/50 contrast-to-saline ratio was used.The affected unit was discarded by the staff post procedure and is not able to be returned.
 
Event or Problem Description
As reported, a 7x40.035 saber percutaneous transluminal angioplasty (pta) balloon ruptured at nominal pressure after being inflated inside of a stent.The balloon ruptured below the rated burst pressure (rbp).The procedure was completed with post-dilatation using a saber.035 pta balloon.No harm or injury occurred to the patient.The product was stored properly per the instructions for use (ifu).The balloon was removed from the hoop, protective cover, and sterile packaging without difficulty.There were no kinks or damages were noted prior to insertion.The device was prepped according to the ifu and maintained negative pressure normally.The intended procedure targeted a lesion in the superficial femoral artery (sfa), which was severely calcified with no vessel tortuosity and 70¿80% stenosis.The device was not used for a chronic total occlusion (cto).No resistance or friction occurred during insertion through the rotating hemostatic valve or while advancing through the vessel or crossing the lesion.The catheter was not in an acute bend and did not kink during use.A 50/50 contrast-to-saline ratio was used.The affected unit was discarded by the staff post procedure and is not able to be returned.
 
Additional Manufacturer Narrative
As reported, a 7x40.035 saber percutaneous transluminal angioplasty (pta) balloon ruptured at nominal pressure after being inflated inside of a stent.The balloon ruptured below the rated burst pressure (rbp).The procedure was completed with post-dilatation using a saber.035 pta balloon.No harm or injury occurred to the patient.The product was stored properly per the instructions for use (ifu).The balloon was removed from the hoop, protective cover, and sterile packaging without difficulty.There were no kinks or damages were noted prior to insertion.The device was prepped according to the ifu and maintained negative pressure normally.The intended procedure targeted a lesion in the superficial femoral artery (sfa), which was severely calcified with no vessel tortuosity and 70¿80% stenosis.The device was not used for a chronic total occlusion (cto).No resistance or friction occurred during insertion through the rotating hemostatic valve or while advancing through the vessel or crossing the lesion.The catheter was not in an acute bend and did not kink during use.A 50/50 contrast-to-saline ratio was used.The product was not returned for analysis.A review of the manufacturing documentation associated with lot 82276443 presented no issues during the manufacturing process that can be related to the reported event.Based on the limited information provided, it is not possible to draw a conclusion about a clinical relationship between the device and the event.Without the return of the device for analysis, the reported customer complaint ¿balloon burst at/below rbp¿ could not be evaluated.Without the return of the device the exact cause of the reported event could not be determined.According to the warnings in the safety information in the instructions for use, ¿prior to angioplasty, the catheter should be examined to verify functionality and integrity, and ensure that its size and shape are suitable for the specific procedure for which it is to be used.Do not use if product damage is suspected or evident.To reduce the potential for vessel damage or the risk of dislodgement of particles it is very important that the inflated diameter of the balloon should approximate the diameter of the vessel just proximal and distal to the lesion.The balloon dimensions are printed on the product label.The compliance table incorporated with the product shows how balloon diameter increases as pressure increases.Do not exceed the rated burst pressure recommended on the label.The rated burst pressure is based on the results of in vitro testing.At least 99.9% of the balloons (with a 95% confidence) will not burst at or below their rated burst pressure.Use of a pressure monitoring device is recommended to prevent over-pressurization.Pressure in excess of the rated burst pressure can cause balloon rupture and potential inability to withdraw the catheter through the introducer sheath.Balloon rupture can cause vessel damage and the need for additional intervention.Use only the recommended balloon inflation medium (a 50/50 mixture by volume of contrast medium and normal saline).Never use air or any gaseous medium to inflate the balloon.¿ based on the information available for review, there is no indication to suggest that the reported failure could be related to the design or manufacturing process of the unit.Therefore, no corrective or preventive action will be taken at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SABER035
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
CORDIS CORPORATION
14201 nw 60 avenue
miami lakes FL 33014
Manufacturer (Section G)
CORDIS US CORP.
14201 nw 60 avenue
miami lakes FL 33014
Manufacturer Contact
karla castro
coyol free zone, building b25, ,  , costa rica
el coyol alajuela bld 25 
7863138372
MDR Report Key23054843
Report Number3007635982-2025-00202
Device Sequence Number9225119
Product Code LIT
UDI-Device Identifier20705032080122
UDI-Public(01)20705032080122(17)260228(10)82276443
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K210626
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/02/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date02/28/2026
Device Catalogue Number48007004R
Device Lot Number82276443
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/21/2025
Supplement Date Received by Manufacturer09/24/2025
Initial Report FDA Received Date09/15/2025
Supplement Report FDA Received Date10/02/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/11/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNKNOWN.
Patient Age74 YR
Patient SexMale
Patient Weight118 KG
-
-