|
A customer reported obtaining a lower than expected, non-reproducible phenytoin (phyt) result from a single patient sample using vitros phyt slide lot 2627-0194-4131 processed on a vitros xt 7600 integrated system.The definitive cause for the event could not be determined historical vitros phyt quality control results using non-vitros biorad quality controls and a vitros phyt within-run precision test indicated vitros phyt slide lot 2627-0194-4131 was performing as intended and did not likely contribute to the event.In addition, continual tracking and trending of complaints has not identified any signals that would point to a potential systemic performance issue with vitros phyt slide lot 2627-0194-4131.A diagnostic vitros crbm within-run precision test, used to assess instrument performance was not performed, therefore, a performance issue with the vitros xt 7600 integrated system cannot be completely ruled out as contributing to the event.However, since historic vitros phyt quality control results and a vitros phyt within-run precision test were acceptable, an instrument-related performance issue did not likely contribute to the event.Finally, the affected sample was not centrifuged according to the collection device manufacture's recommendations, therefore, improper pre-analytical sample handling cannot be ruled out as a contributing factor.It is likely cellular debris, due to poor sample preparation, was present in the affected sample, although this could not be confirmed.
|