H11: additional manufacturer narrative: the investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
|
|
H11: additional manufacturer narrative: updated sections g3, h6 (type of investigation, investigation findings, investigation conclusions).Regurgitation is a common manifestation of bioprosthetic valve and valved conduit dysfunction.Common intraoperative factors that may cause or contribute to acute post-procedural regurgitation include device distortion; device interaction with patient anatomy or other devices; leaflet damage, and incorrectly sized valve seated.Non-structural valve dysfunction (nsvd), endocarditis and/or thrombosis may also cause or contribute to acute post-procedural regurgitation.Complaints reported with regurgitation as the primary complaint code are not typically attributed to manufacturing-related nonconformances.There is no evidence to suggest a manufacturing defect caused or contributed to the reported event.Based on the information available, the most likely root cause is patient factors, including chronic kidney disease.The instructions for use (ifu) have been reviewed, and no inadequacies have been identified regarding warnings, contraindications, and the directions/conditions for the successful use of the device.The reported type of event is included in the ifu.The subject device is not available for evaluation as it remains implanted in the patient.The device history record (dhr) review was completed, and this device passed all manufacturing and sterilization inspections prior to release for distribution.There were no issues identified that would have impacted this event.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
|