MAUDE Adverse Event Report: ZIMMER BIOMET, INC. COMP RVRS 25MM BSPLT HA+ADPTR; SHOULDER PROSTHESIS, REVERSE CONFIGURATION/EXTREMITIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. COMP RVRS 25MM BSPLT HA+ADPTR; SHOULDER PROSTHESIS, REVERSE CONFIGURATION/EXTREMITIES Back to Search Results
Catalog Number 010000589
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Codes Bone Fracture(s) (1870); Muscle Weakness/Atrophy (1967); Pain (1994); Loss of Range of Motion (2032); Nonunion/Delayed-union Bone Fracture (2369); Joint Dislocation (2374); Osteolysis (2377); Unequal Limb Length (4534); Bone Notching (4890)
Date of Event 04/03/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).D10: medical products: item#: 115310, comp rvrs shldr glnsp std 36mm; lot#: 668560.Item#: 00435003600, 36mm vit e liner +0mm; lot#: 64375020.Item#: 115395, comp rvs cntrl 6.5x25mm st/rst; lot#: 916180.Item#: 180551, comp lk scr 3.5hex 4.75x20 st; lot#: 428110.Item#: 180553, comp lk scr 3.5hex 4.75x30 st; lot#: 457940.Item#: unknown, unknown tm rev humeral component; lot#: unknown.G2: foreign: japan.G2: literature: yuki yoshida, takeshi ikegami, jos case reports, https://doi.Org/10.1016/j.Joscr.2025.02.001.H3: customer has indicated that the product will not be returned to zimmer biomet for investigation, as the product location is unknown.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that the patient underwent an initial reverse total shoulder on an approximately four (4) years and seven (7) months ago.Subsequently, the patient underwent a revision surgery approximately one (1) year and five (5) years ago due to increased pain and loss of range of motion due to the implant disassociating with clavicular nonunion and scapular malalignment.
 
Additional Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional and corrected information.The following sections were updated: b4; b5; d2; g2; g3; g6; h1; h2; h3; h6.The following section was corrected: d10.D10: medical products: item#: 00434901213; 12mm, 130mm length humeral stem; lot#: 64759372.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history record(s) identified no deviations or anomalies during manufacturing related to the reported event.Medical records were provided and reviewed by a health care professional.Review of the available records identified the following: a reverse shoulder arthroplasty was placed in the patient on (b)(6) 2021 at the age of 73.The final routine checkup in 2022 (month unknown) noted, bone resorption of the greater tubercle and scapular notching.At 75 years of age, she began experiencing frequent falls due to muscle weakness eventually requiring admission to a care facility.After admission, there was no documented history of falls or trauma.However, a palpable mass was observed in her right shoulder.Although she experienced pain during shoulder movement while receiving care, the condition was initially monitored without intervention.Her shoulder pain during caregiving gradually worsened, and she was referred to hospital at 77 years of age.On (b)(6) 2024, the patient presented to the emergency department complaining of shoulder pain.Radiography and computed tomography revealed that the palpable mass was the dislocated glenosphere at the anterior surgical site, along with glenosphere dissociation.Revision occurred on (b)(6) 2024 when the glenosphere, liner, baseplate and screws were revised and the humeral stem was retained.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMP RVRS 25MM BSPLT HA+ADPTR
Common Device Name
SHOULDER PROSTHESIS, REVERSE CONFIGURATION/EXTREMITIES
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key23057752
Report Number0001825034-2025-02905
Device Sequence Number9347160
Product Code PHX
UDI-Device Identifier00880304532465
UDI-Public(01)00880304532465(17)301130(10)919300
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K193373
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Literature,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/23/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue Number010000589
Device Lot Number919300
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/18/2025
Supplement Date Received by Manufacturer09/30/2025
Initial Report FDA Received Date09/15/2025
Supplement Report FDA Received Date10/24/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/30/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age77 YR
Patient SexFemale
-
-